SAFETY
2-2 30 Sep 13
DE-7200VFDCentrifuge
WARNING! ELECTRICAL CONNECTIONS MUST BE MADE IN ACCORDANCE
WITH ALL APPLICAB
LE NATIONAL AND LOCAL CODES. FAILURE TO
COMPLY MAY RESULT IN AN UNSAFE CONDITION THAT COULD INJURE
PERSONNEL OR DAMAGE EQUIPMENT. ENSURE THAT ALL ELECTRICAL
AND CONDUIT CONNECTIONS ARE SECURE.
Equipment Handling
WARNING! TO ENSURE PROPER BALANCE AND ORI
ENTATION WHEN UNIT
IS RAISED AND PREVENT DAMAGE TO COMPONENTS, ATTACH LIFTING
SLING ONLY AT DESIGNATED LIFT POINTS. DO NOT ATTEMPT LIFTING BY
ATTACHMENT TO MOTOR OR ANY OTHER LOCATION.
WARNING! BE SURE THAT HANDLING DEVICES HAVE SUFFICIENT LIFTING
CAPAC
ITY TO SAFELY HANDLE THE WEIGHT OF THE EQUIPMENT.
WARNING! WHEN USING AN OVERHEAD LIFTING DEVICE, USE ALL FOUR
LIFTING POINTS PROVIDED.
WARNING! DO NOT REMOVE SHIPPING BRACKETS UNTIL EQUIPMENT HAS
BEEN POSITIONED AT FINAL INSTALLATION SITE.
WARNIN
G! WHILE OPENING AND CLOSING THE CENTRIFUGE COVER,
PERSONNEL MUST ENSURE THAT ALL BODY LIMBS AND ANY OTHER
ITEMS ARE CLEAR OF THE COVER GAP. TWO HANDS MUST BE USED,
WITH BOTH HANDS ON THE COVER HANDLES AT ALL TIMES. DERRICK
RECOMMENDS HAVING A HOIST AVAILA
BLE IF ADDITIONAL ASSISTANCE
IS DESIRED DURING THE OPENING/CLOSING PROCEDURE.
Operation
WARNING! ALL OPERATING AND MAINTENANCE PERSONNEL MUST READ
AND UNDERSTAND ALL SAFETY INFORMATION IN THIS MANUAL BEFORE
WORKING WITH THE EQUIPMENT.
WARNING! BE SUR
E THAT TOP COVER IS CLOSED AND SECURED AND
ALL PERSONNEL ARE CLEAR BEFORE STARTING MACHINE.
WARNING!
BEFORE STARTING CENTRIFUGE, BE SURE THAT ALL
SHIPPING BRACKETS HAVE BEEN REMOVED AND BEARING PILLOW
BLOCKS ARE PROPERLY TIGHTENED.
WARNING! ALWAYS AL
LOW MACHINE TO COAST TO A COMPLETE STOP
BEFORE OPENING
TOP COVER OR REMOVING GUARDS.
WARNING! DO NOT OPERATE CENTRIFUGE IF EXCESSIVE NOISE OR
VIBRATION DEVELOPS. ALWAYS CONFIRM THAT VIBRATION SWITCH AND
OTHER SAFETY DEVICES ARE FUNCTIONAL.