INSTALLATION
01 Jan 14 3-9
DE-7200VFDCentrifuge
Liquid Discharge
The liquid discharge collector may be bolted directly to the centrifuge case; however, the piping
connection should have a flexible joint. When fabricating the discharge piping, follow the
dimensional and piping requirements shown in Figure 3-6. Since liquid exiting the centrifuge is not
pressurized, the piping must be pitched to facilitate gravity drainage.
Solid Discharge Chute
A three-sided solid discharge extension may be bolted to the discharge flange. The chute may not
be connected to any other component or structure; it must be supported or suspended
independently of the centrifuge. The chute should be designed with a sufficiently steep angle to
facilitate self cleaning. If this is not possible, a means should be included to continually remove
solids from beneath the centrifuge. A curtain or belting should surround three sides of the
discharge chute to contain spray and mist.
EQUIPMENT LEVELING
The centrifuge must be properly leveled for satisfactory operation. The equipment must be leveled
along the length and width of the unit (Figure 3-7). A 2-foot or torpedo level is recommended.
Non-compressible shims should be used as required to level the machine.
Figure 3-7. Centrifuge Leveling