MAINTENANCE
5-34 31 Dec 13
DE-7200VFDCentrifuge
ALARM AND FAULT MESSAGES
Cause Corrective Action
F9012 - External Communication Loss, Switched to Local Control
Profibus not working, remote control is
disabled and control has automatically
switched to local
Check and correct any defects in Profibus
connections, and attempt to re-start in Remote
operating mode.
F9021 - Signal Loss - Bowl Speed Sensor Intrinsic Barrier
Minimum analog signal from
transmitter/intrinsic barrier not received
by PLC
Check all connections at transmitter/intrinsic barrier;
replace transmitter if all connections are secure.
F9022 - Signal Loss - Bowl Speed Sensor or Cable
Incorrect or loose connection at speed
sensor or defective speed sensor
Check sensor connections; replace sensor if
defective.
Defective speed sensor
transmitter/intrinsic barrier
Check and replace transmitter/intrinsic barrier, if
defective.
Pump VFD fault Identify fault from operator control panel. Perform
indicated corrective action.
F9031 - Signal Loss - Clutch Speed Sensor Intrinsic Barrier
Minimum analog signal from
transmitter/intrinsic barrier not received
by PLC
Check all connections at transmitter/intrinsic barrier;
replace transmitter if all connections are secure.
F9032 - Signal Loss - Bowl Speed Sensor or Cable
Incorrect or loose connection at speed
sensor or defective speed sensor
Check sensor connections; replace sensor if
defective.
Defective speed sensor
transmitter/intrinsic barrier
Check and replace transmitter/intrinsic barrier, if
defective.
Pump VFD fault Identify fault from operator control panel. Perform
indicated corrective action.
F9041 - Signal Loss - Liquid End Bearing Temp Intrinsic Barrier
Minimum analog signal from
transmitter/intrinsic barrier not received
by PLC
Check all connections at transmitter/intrinsic barrier;
replace transmitter if all connections are secure.
F9051 - Signal Loss - Solid End Bearing Temp Intrinsic Barrier
Minimum analog signal from
transmitter/intrinsic barrier not received
by PLC
Check all connections at transmitter/intrinsic barrier;
replace transmitter if all connections are secure.